Latiss Master Calibrations Generation Procedure¶
Overview¶
This procedure describes how to execute the SAL Script to produce master calibrations for LATISS on the ScriptQueue from the LSST Operations and Visualization Enviroment (LOVE) at the Summit operational environment.
The script will have the option to:
- command LSSTComCam to take a number of calibrations frames:
- biases
- darks
- flats
- call the corresponding Rubin Science Pipelines calibration generation pipetask to produce biases, darks, flats, defects and Photon Transfer Curves (PTCs) via the OCS-Controlled Pipeline System (OCPS),
- verify the resulting calibration (see the package cp_verify and DMTN-101),
- certify the resulting calibration with a given range of validity dates, if a minimum number of verification tests passed.
For at least one type of test (as defined in DMTN-101), if the majority of tests fail in the majority of detectors and the majority of exposures, then the script will terminate by raising a RuntimeError after calculating the verification statistics, and the calibration will not be certified. The configuration parameters number_verification_tests_threshold_bias, number_verification_tests_threshold_dark, and number_verification_tests_threshold_flat will be used to define thresholds to decide whether the calibration will pass verification and should be certified or not. Currently, verification is only implemented for BIAS
, DARK
, and FLAT
calibration types. If the configuration parameters do_defects and do_ptc are set to True
, verification will be skipped for the DEFECTS
and PTC
calibrations and they will be automatically certified.
The script currently has the option (via the script_mode parameter in the configuration options) to:
- take only biases,
- take biases and darks, and
- take biases, darks, and flats.
These options are constrained by the fact that one calibration depends on the existence of the previous one (i.e., to build a dark, a bias is necessary, and to build a flat, a dark and a bias are necessary).
If desired, defects can be constructed from darks and flats, and a PTC per detector per amplifier constructed from the flats. Note that the PTC assumes that a sequence of flat pairs has been taken, each pair taken at the same exposure time. In both cases, script_mode must be set to BIAS_DARK_FLAT
.
For more information about calibrations production (including verification and certification), please consult the Constructing Calibrations documentation.
Prerequisites¶
- You should be logged into the LSST Operations and Visualization Enviroment (LOVE) at the Summit operational environment.
- The script assumes (and checks) that the
LATISS
andOCPS
components are allENABLED
, and that the latter has beenENABLED
with the configuration ofLATISS
.
- The instrument and the
OCPS
can be enabled with the following procedures: - If you plan to take flat fields as well, make sure the Auxiliary Telescope is prepared to do so:
Post-Condition¶
- A (daily) master calibration image per detector will be certified in a butler
CALIBRATION
collection.
Procedure Steps¶
Once you are logged into LOVE, click on the ATQueue
panel, as circled on the left side of the figure below (for reference, MTQueue
to launch LSSTComCam
scripts is circled on the right):
Load the Script¶
After clicking on the ATQueue
panel, search for the script auxtel/make_latiss_calibrations.py
under AVAILABLE SCRIPTS
on the left, as shown in the figure below:
Load the script by clicking on the button in front of the name of the script that has a triangle.
Enter configuration parameters¶
After loading the script, a window that contains two sections, SCHEMA
(top) and CONFIG
(bottom), will appear. The former will show the available configuration options (and the default values of some of them) that should be entered in the latter. The configuration options are as follows:
- script_mode: Currently, the script can be run in three modes, in which it will produce only biases (
BIAS
), biases and darks (BIAS_DARK
), or biases, darks, and flats (BIAS_DARK_FLAT
). Default:BIAS_DARK_FLAT
- n_bias: number of bias frames to be taken. Default: 1
- n_dark: number of dark frames to be taken. Default: 1
- exp_times_dark: The exposure time of each dark image (sec). If a single value, then the same exposure time is used for each exposure. Default: 0
- n_flat: number of flat frames to be taken. Default: 1
- exp_times_flat: The exposure time of each flat image (sec). If a single value, then the same exposure time is used for each exposure. Default: 0
- detectors: Detector IDs, e.g.,
(0,1,2,3,4,5,6,7,8)
for all LATISS CCDs. Default: “(0,1,2,3,4,5,6,7,8)” - do_verify: Should the master calibrations be verified? (c.f.,
cp_verify
). Default: True - number_verification_tests_threshold_bias: Minimum number of verification tests per detector per exposure per test type that should pass to certify the bias master calibration. Default: 8
- number_verification_tests_threshold_dark: Minimum number of verification tests per detector per exposure per test type that should pass to certify the dark master calibration. Default: 16
- number_verification_tests_threshold_flat: Minimum number of verification tests per detector per exposure per test type that should pass to certify the flat master calibration. Default: 8
- config_options_bias: Options to be passed to the command-line bias pipetask. They will overwrite the values in
cpBias.yaml
. Default: “-c isr:doDefect=False -c isr:doLinearize=False -c isr:doCrosstalk=False -c isr:overscan.fitType=’MEDIAN_PER_ROW’” - config_options_dark: Options to be passed to the command-line dark pipetask. They will overwrite the values in
cpDark.yaml
. Default: “-c isr:doDefect=False -c isr:doLinearize=False -c isr:doCrosstalk=False” - config_options_flat: Options to be passed to the command-line flat pipetask. They will overwrite the values in
cpFlat.yaml
. Default: “-c isr:doDefect=False -c isr:doLinearize=False -c isr:doCrosstalk=False -c cpFlatMeasure:doVignette=False “ - do_defects: Should defects be built using darks and flats?. script_mode must be
BIAS_DARK_FLAT
.Default: False - config_options_defects: Options to be passed to the command-line defects pipetask. They will overwrite the values in
findDefects.yaml
. Default: “-c isr:doDefect=False “ - do_ptc: Should a Photon Transfer Curve be constructed from the flats taken?
script_mode
must beBIAS_DARK_FLAT
. Default: False - config_options_ptc: Options to be passed to the command-line PTC pipetask. They will overwrite the values in
measurePhotonTransferCurve.yaml
. Default: “-c ptcSolve:ptcFitType=EXPAPPROXIMATION -c isr:doCrosstalk=False “ - input_collections_bias: List of additional (the
OCPS
already addsLATISS/raw/all
as a default) comma-separated input collections for the bias pipetask. The pipetask is called via theOCPS
after enabling it with theLATISS
configuration. Default: “LATISS/calib”. - input_collections_verify_bias: Additional comma-separated input collections to pass to the verify (bias) pipetask. Default: “LATISS/calib”.
- input_collections_dark: Additional comma-separarted input collections to pass to the dark pipetask. Default: “LATISS/calib”
- input_collections_verify_dark: Additional comma-separated input collections to pass to the verify (dark) pipetask. Default: “LATISS/calib”
- input_collections_flat: Additional comma-separated input collections to pass to the flat pipetask. Default: “LATISS/calib”
- input_collections_verify_flat: Additional comma-separated input collections to pass to the verify (flat) pipetask. Default: “LATISS/calib”
- input_collections_defects: Additional comma-separated input collections to pass to the defects pipetask. Default: “LATISS/calib”
- input_collections_ptc: Additional comma-separated input collections to pass to the Photon Transfer Curve pipetask. Default: “LATISS/calib”
- calib_collection:
CALIBRATION
collection where the calibrations will be certified into, for example,LATISS/calib/u/plazas/YYYYMMMDD.test
. Default: “LATISS/calib/daily”. - repo: Butler repository. Default:
/repo/LATISS
. - n_processes: Number of processes that the pipetasks will use. Default: 8
- certify_calib_begin_date: The beginning date for the validity range of the certified calibration. For example,
2021-07-15
. Default: “1950-01-01” - certify_calib_end_date: The end date for the validity range of the certified calibration. For example,
2021-07-16
. Default: “2050-01-01” - oods_timeout: Timeout value, in seconds, for the Observatory Operations Data Service (
OODS
). Default: 120
An example set of configuration parameters is as follows:
n_bias: 6
n_dark: 6
exp_times_dark: [5, 5, 5, 10, 15, 20]
n_flat: 14
exp_times_flat: [0.1, 0.1, 0.35, 0.35, 0.6, 0.6, 0.8, 0.8, 1.0, 1.0, 1.35, 1.35, 1.6, 1.6]
detectors: "(0)"
calib_collection: "LATISS/calib/u/plazas/daily.2021SEP13.test1"
input_collections_verify_bias: "u/czw/DM-28920/calib.20210720,LATISS/calib"
input_collections_verify_dark: "u/czw/DM-28920/calib.20210720,LATISS/calib"
input_collections_verify_flat: "u/czw/DM-28920/calib.20210720,LATISS/calib"
script_mode: "BIAS_DARK_FLAT"
do_defects: True
do_ptc: True
Launch the script¶
When the configuration options have been entered and the script is ready to be launched, click on the ADD
button in the lower right of the screen (refer to image above).
Accessing the calibrations¶
The certified master calibrations will be available via the collection specified by the calib_collection parameter. They could be retrieved from a notebook for manipulation and visualization:
import lsst.daf.butler as dB
butler = dB.Butler("/repo/LATISS", collections=["LATISS/calib/daily.2021SEP13.test1"])
detector = (0, 1, 2, 3, 4, 5, 6, 7, 8)
exposure = [bias1ID, bias2ID] # e.g., [2021071500001, 2021071500002]
# For detector "0":
bias = butler.get('bias', detector=detector[0], exposure=exposure[0], instrument='LATISS')
dark = butler.get('dark', detector=detector[0], exposure=exposure[0], instrument='LATISS')
flat = butler.get('flat', detector=detector[0], exposure=exposure[0], instrument='LATISS')
defects = butler.get('defects', detector=detector[0], exposure=exposure[0], instrument='LATISS')
ptc = butler.get('ptc', detector=detector[0], exposure=exposure[0], instrument='LATISS')
In addition, the statistics produced by the verification step can be analized by looking into the examples
folder in cp_verify
.
Troubleshooting¶
After checking the configuration options and theLOVE
error messages, the file/scratch/uws/${jobId}/outs/ocps.log
will contain additional technical information on which pipetask failed, if any.{jobId}
is returned by the OCPS and can be retrieved from theLOVE
output messages.
Contact Personnel¶
This procedure was last modified on Nov 10, 2021.
This procedure was written by Andrés A. Plazas Malagón. The following are contributors: none.